Physical Specifications |
|||
Dimension and Weight |
Dimension |
36cm (W)*23cm (D)*12cm (H) |
|
Weight |
5.3kg(excluding accessories) |
||
Operation Environment |
Power requirements |
AC100-240V , 50/60Hz |
|
Power |
100VA |
||
Temperature |
0-40ºC |
||
Humidity |
0-93% non-condensing |
||
Fetal/Maternal Performance Specifications |
|||
Performance Specifications |
Display |
12.1 inch (1024×768 pixel) color TFT -LCD |
|
Indicator |
Visual on-display and integrated, alarm light, audible alarm |
||
Battery |
Internal rechargeable Li-ion battery |
||
Alarm |
Visual on-display and integrated, alarm light, audible alarm |
||
Memory |
24-hours memory for fetal traces 150 groups patient data |
||
Printer |
America standard/International standard Built-in thermal printer (110mm, 150mm for option) 1/2/3cm/min real-time printing speed, fast print speed (stored traces) |
||
Paper |
Z-fold, thermosensitive |
||
Record Message |
ID, name, age, date, time, print speed etc |
||
CMS |
Support connecting to central monitoring station through LAN (TCP/IP) or wireless connection |
Fetal/Maternal Technical Specifications |
|||
Ultrasound (FHR) |
Measurement range |
US |
30 to 240 bpm |
Resolution |
Display |
1bpm |
|
Printer |
1bpm |
||
jitter@200 bpm |
≤3bpm |
||
Display update rate |
0.25second |
||
Beat to bea change(max.) for ultrasound |
28bpm |
||
US frequency |
1MHz±5%Hz |
||
TOCO |
Measurement method |
Strain gauge sensor element |
|
Sensitivity |
1 unit =2.5g |
||
Resolution |
1 unit |
||
Measurement range |
500 units |
||
Baseline setting |
0, 5,10, 15, 20 can be selective |
||
Update rate |
0.25 second |
||
Auto offset correction |
3 second after connecting the transducer |
||
Auto zero adjust
|
TOCO value is set to zero negative following a negative Measure value for 5 seconds |
||
Auto Fetal Movement (AFM) |
Technique |
Pulsed doppler ultrasound |
|
Range |
0-100% |
||
Resolution |
1% |
||
Automatic fetal movement actograph |
|||
Manual fetal movement marker |
Maternal NIBP(MNIBP) |
Operation mode |
Manual/automatic/ STAT |
|
Measurement unit |
mmHg / kPa selectable |
||
Test interval |
0-480 minutes |
||
Accuracy |
Max.Std.deviation:8 mmHg(1.1kPa) |
||
Max. mean Error: ±5mmHg(±0.7kPa) |
|||
Measurement range |
Systolic |
30 to 270 mmHg(4 to 36 kPa) |
|
Diastolic |
10 to 245 mmHg(1.5 to 32kPa) |
||
Mean |
20 to 255 mmHg(2.5 to 34kPa) |
||
Pulse rate range |
20~250bpm |
||
Over-pressure protection |
Yes |
||
Maternal ECG (MECG) |
Heart rate range |
0~300bpm |
|
Resolution |
1bpm |
||
Accuracy |
±1bpm/ ±1% |
||
Leads |
I, II, III, avL, avR, avF, V |
||
Sweep speed |
12.5mm/s, 25mm/s, 50mm/s |
||
Gain selection |
x 0.25, x0.5 ,x1, x2 |
||
Operation mode |
diagnostic, monitor, surgery |
||
Maternal SpO2 (MSpO2) |
Measurement range |
0%-100% |
|
Resolution |
1% |
||
Accuracy |
±2% |
||
Pulse rate range accuracy |
±3bpm |
||
Maternal Temperature (MTEMP) |
Technique |
Thermistor probe CYSI 400 series compatible |
|
Measurement range |
0ºC-50ºC |
||
Accuracy |
±0.1ºC |