Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Infusion rate (ml/h) | 1~400 ml/hr in 1 ml increments |
Mean flow velocity accuracy | ±10% of selected flow rate |
Dosage limit(ml) | 0-9999 in 1 ml increment |
Meanvolume accuracy | ±10% |
Loading speed andmeanaccuracy (ml/h) | 1000±10% |
Cleaning speed andmeanspeedaccuracy(ml/h) | 1000±10% |
Max volume flushandmeanspeed accuracy(ml) | 40±10% |
Mx feeding pump pressure | ≥0.1 MPa |
Occlusion alarm level (pressure) | 50KPa±10KPa |
The longest occlusion triggering time of this device at minimum flow rate | 43min |
The shortest occlusion triggering time of this device at intermediate flow rate | 85s |
Medium speed running the pills from blocking the alarm threshold | 0.79ml |
Recoverable Alarm | 1min50s-2min |
Pause overtime alarm time | 1min50s-2min |
Heating temperature | Infusion flow rate ≥20ml / h, 375px distance away from the heater outlet at 25-38 ºC |
Use ofinfusionset | See recommended infusion sets list |
Air-in-line sensitivity | Detestability of single size of air bubble ≥50uL |
Alarm | Occlusion alarm, Over heatedalarm, battery depleted alarm |
Alarms | Air in line,Prime, Flush, low power, failurecodes (ERxx), pause overtime, powerstate, infusion mode, work state, heaterstate. |
Noise | Measured at the highest flow rate on wooden table, the volume is lower than 50 dB at a distance of 1 meter |
Power | Ac: 100V-240V 50 Hz /60Hz |
Built inbattery | A fully charged internal DC11.1V battery can continuously run the pump for approximately 5 hours on intermediate rate |
Power consumption | ≤40VA |
Power output | ≤20VA |
The rated output voltage | 15V |
Weight | 2.0kg |
Heater | W-E01 ,cable length is 1.5m,Shield |
Essential performance | This product is suitable for patients with enteral nutrition infusion, the host cancontrol the infusion rate. With alarmfunction, with bubble detection piping and piping blockage detection. |
Classification | type BF,class ΙΙ continuous volumetric feeding pump with internal battery; IPX1 (sealed drip-proof device) |
Dimensions | 155mm(L)×170mm(W)×180mm(H) |
Accessories | 1 user manual, 1 power cable, 1 heater |