Customization: | Available |
---|---|
Type: | Baby Care Equipment |
Function: | Infant Treatment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
B-010 Standard Hospital Use Baby/Infant Incubator
Features:
1.Microprocessor based servo controlled temperature system |
2.Air temperature,skin temperature,humidity,timer and heat power separately displayed by LED |
3.Alarms functions : Power failure; over temperature; temperature deviation; temperature sensor failure;fan failure |
4.Silence/reset button for alarms |
5.Integral timer to record therapy time |
6.Humidity is adjustable continuously |
7.>37°C temperature setting |
8.A keypad lock prevent inadvertent changes the setting |
9.Trough for pulling and pushing ,which is convenient to add water and wash |
10.Fan with low noise and long life-span |
11.Temperature deviation and over temperature could be amended in front panel directly |
12.A second thermal cut-off function for more safety |
13.Skin temperature sensor failure protect function avoid over temperature |
14.The infant bed inclination angle is adjustable |
15.six operating windows,two sides operating windows can be swiveled |
16.Heater is electrically cut-out if air temperature inside incubator exceed 40°C |
17.big storage cabinet |
Standard configuration
Main body(including the transparent hood,control system,infant bed,water trough),
skin temperature sensor,I.V.pole, mattress, height adjustable ;cabinet, castors
Optional parts:
Phototherapy up side unit |
Phototherapy from up and down sides |
RS232(485) connector |
Tray/pole |
Technical data:
Power supply | AC 110/220V 60/50Hz |
Power input | ≤420VA |
Operating conditions |
Environment temperature: 20°C~30°C Environment relative humidity: 30%~75%RH Environment air velocity of flow: <0.3m/s |
Air Temperature control range | 25°C~38°C |
Skin temperature control range | 32°C~38.5°C |
Skin temperature sensor accuracy | +/-0.3°C |
Temperature fluctuation | +0.5°C |
Uniformity of mattress temperature | ≤0.8°C |
Warm-up time | ≤45min |
Internal noise level | ≤55dB |
Humidity display range | 10%-95%RH |
Humidity accuracy | ≤10%RH |
Infant bed tilt angle | +10º |
Mattress size | 67cm(L)×54cm(W) |
Trough capacity | 1200ml |
Air filter | 0.5um |
Transport and storage
Each unit is packed in two wooden crate
Size (control unit and cabinet):
120X 70X 96CM
N. W.: 80KG G. W.: 110KG
Hood: 97X 56X 63CM
N. W.: 15KG G. W.: 40KG
Environment Temp | -10°C--+70°C |
Environment Relative Humidity | ≤80% RH |
Atmospheric Pressure | 500~1060Hpa |